NIHON KOHDEN CORPORATION BSM-1733A; VITAL SIGNS MONITOR Back to Search Results
Model Number BSM-1733A
Device Problems No Device Output (1435); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2023
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that the spo2 stopped working on the bedside monitor (bsm).Qa and tech support contacted the customer for information on the unit but they have been unresponsive.It was unclear whether spo2 stopped working in the midst of monitoring a patient; or whether they were already monitoring other vitals for the patient on the unit, and discovered this issue when they tried to use the spo2 and found it did not work.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.The following fields contains no information (ni), as attempts to obtain information were made, but the information was not provided.A2: a6, b6: b7.D10: concomitant medical device.Attempt #: 1.(b)(6) 2024 emailed customer via microsoft outlook for all items under the no information section.No reply was received.Attempt #: 2.(b)(6) 2024 emailed customer via microsoft outlook for all items under the no information section.No reply was received.Attempt # 3.01/09/2024 emailed customer via microsoft outlook for all items under the no information section.No reply was received.
 
Event Description
The biomedical engineer (bme) reported that the spo2 stopped working on the bedside monitor (bsm).Qa and tech support contacted the customer for information on the unit but they have been unresponsive.It was unclear whether spo2 stopped working in the midst of monitoring a patient; or whether they were already monitoring other vitals for the patient on the unit, and discovered this issue when they tried to use the spo2 and found it did not work.No patient harm was reported.
 
Event Description
The biomedical engineer (bme) reported that the bedside monitor (bsm) stopped giving spo2 readings.It was unclear whether the spo2 stopped working while in patient use or was discovered when trying to first use it.No patient harm was reported.
 
Manufacturer Narrative
Details of complaint: the biomedical engineer (bme) reported that the bedside monitor (bsm) stopped giving spo2 readings.It was unclear whether the spo2 stopped working while in patient use or was discovered when trying to first use it.No patient harm was reported.Investigation summary: the reported device, (bsm-1733a, sn: (b)(6)) was returned and evaluated.During evaluation of the device, nihon kohden repair center (nk rc) confirmed the reported issue.The spo2 computer board, a/msx-2013 (spo2 brd masimo), had an electrical failure, and was replaced.After the part was replaced, the device passed all post repair testing and evaluation, meeting specifications in all areas.The device was then shipped back to the customer, resolving the issue.The root cause was attributed to a component failure, possibly due to a power issue, such as a power outage or sudden surge in power at the facility, leading to damage to the sensitive component.Once the issue was resolved, no further issues related to this incident have been reported since.A review of the device history showed that this was the only issue found for this device at this facility, in the past three (3) years.A significant trend that would warrant any action and/or any corrective action has not been observed.Trending will continue to be monitored.The following fields contains no information (ni), as attempts to obtain the information were made, but not provided: attempt #1 01/02/2024 emailed customer via microsoft outlook for all items under the no information section.No reply was received.Attempt #2 01/04/2024 emailed customer via microsoft outlook for all items under the no information section.No reply was received.Attempt #3 01/09/2024 emailed customer via microsoft outlook for all items under the no information section.No reply was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BSM-1733A
Type of Device
VITAL SIGNS MONITOR
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key18585192
MDR Text Key334779889
Report Number8030229-2024-04062
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921111833
UDI-Public4931921111833
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBSM-1733A
Device Catalogue NumberBSM-1733A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2023
Initial Date FDA Received01/25/2024
Supplement Dates Manufacturer Received04/10/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-