U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX500H11C
Device Problem Degraded (1153)
Patient Problems Pneumonia (2011); Choking (2464); Cough (4457)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device not returned to the manufacturer.
 
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging water entered the lungs resulting in pneumonia.The patient was seen by a the doctor who took x-rays and the patient was diagnosed with pneumonia.The patient's current condition was described as okay, but the patient had a cough.The device settings at the time of the alleged incident was 12cm (centimeters of water pressure).The cleaning method the patient used was rinsing the device with water and mild detergent, then rinsed the device again and dried completely.This method of cleaning was performed daily.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer previously reported information in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging water entered the lungs resulting in pneumonia.The patient was seen by a the doctor who took x-rays and the patient was diagnosed with pneumonia.The patient's current condition was described as okay, but the patient had a cough.The device settings at the time of the alleged incident was 12cm (centimeters of water pressure).The cleaning method the patient used was rinsing the device with water and mild detergent, then rinsed the device again and dried completely.This method of cleaning was performed daily.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.Correction/update: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging the water was draining too quickly and was causing the patient to choke at times.The patient reported water entered the lungs resulting in pneumonia.The patient was seen by a the doctor who took x-rays and the patient was diagnosed with pneumonia.The patient's current condition was described as okay, but the patient had a cough.The device settings at the time of the alleged incident was 12cm (centimeters of water pressure).The cleaning method the patient used was rinsing the device with water and mild detergent, then rinsed the device again and dried completely.This method of cleaning was performed daily.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.Please note (the following updates were performed): section b: adverse event/product problem: changed to' product problem'.Removed selection for outcomes attributed to ae since complaint was assessed as non-serious.Section h: type of reported complaint: changed to 'product problem'.Section h: patient outcome code grid: additional patient outcome code added - choking.Section h: health impact grid: changed to 'no health consequences or impact'.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18586332
MDR Text Key333837836
Report Number2518422-2024-04535
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500H11C
Device Catalogue NumberDSX500H11C
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/27/2023
Initial Date FDA Received01/26/2024
Supplement Dates Manufacturer Received12/27/2023
Supplement Dates FDA Received01/26/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1974-2021
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient SexFemale
-
-