It was reported that on (b)(6) 2024, a mitraclip procedure was performed to treat functional mitral regurgitation (mr) grade 3+.An xtw (lot: 30731r1120) was chosen for implantation.During the gripper alignment above the valve, the posterior gripper failed to drop.Troubleshooting was attempted but was unsuccessful.The xtw was removed and a replacement mitraclip was used to complete the procedure, reducing mr to grade 1+.Outside of the patient, the posterior gripper on the xtw still was not functioning.There was no patient consequences and no clinically significant delay.No additional information was provided.
|
All available information was investigated, and the reported difficult to open or close (gripper actuation - single) could not be confirmed via returned device analysis.Additionally, the gripper arm cover (non-tactile) was observed to be bulky and frayed.One gripper lever assembly tab was observed to be broken.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on available information, a cause of the reported difficult to open or close (gripper actuation - single) could not be determined as the issue was not identified in device analysis.A cause of the observed bulky and frayed gripper arm cover could not be determined.The observed broken gripper lever assembly tab appears to be related to procedural handling.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|