Check of the device history record confirmed that savvywire lot number osw-0093 was released per specifications.No non-conformity nor deviation were associated with this lot.Visual inspection of the returned savvywire identified that the corewire of the tip broke approximately 20 mm from the distal tip.The coil has been uncoiled from the corewire over a length of approximately 8 cm.The provided angio images were reviewed but unfortunately did not capture the event.Opsens reached out to the physician for additional information, he informed us that the case initially went smoothly, valve was successfully implanted using savvywire and delivery system removed.Upon removal, no guiding catheter was used over savvywire, a significant resistance was noticed during the withdrawal attempts.Subsequently, force was applied to exist the femoral site.Savvywire's instruction for use clearly states the following warning: - "the savvywire should only be introduced in, and pulled out from, the ventricle through a catheter already positioned in the ventricle.A diagnostic pigtail catheter is recommended." the following precautions are mentioned in the ifu as well: -" do not grasp the tip.The stiffness of the guidewire may cause it to uncoil unexpectedly if pulled out or manipulated abruptly.- never advance, pull or torque a savvywire against resistance without first determining the reason for the resistance under fluoroscopy.".
|