MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
Model Number ETBF2816C145EE |
Device Problem
Difficult to Remove (1528)
|
Patient Problems
Perforation of Vessels (2135); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/24/2024 |
Event Type
malfunction
|
Event Description
|
An endurant ii stent graft (etbf2816c145ee) was implanted during the endovascular treatment of an abdominal aortic aneurysm. it was reported during the index procedure, the endurant stent graft was successfully implanted, however issues were encountered during removal of the delivery system.Despite the physician correctly performing the pull-back maneuver, the trigger of the delivery system became stuck.The delivery system had to be removed without closing it properly, resulting in the physician having to fix the femoral artery.Per the physician the cause of the delivery systems trigger/ removal issues was due to a prothesis fault.No additional clinical sequelae were provided, and the patient is fine.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
B5; additional information received : it was confirmed there was no damage noted to the delivery system prior to use.There was no calcification or thrombus noted at the access vessel.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Film analysis: the reported difficulty in removing the delivery system could not be confirmed based on the pre-implant films provided; therefore, the cause of the event could not be determined.Procedural angiogram videos showing the device removal after deployment were not available for a comprehensive evaluation of the event.Analysis of the pre-implant images did not reveal any anatomical characteristics that could have contributed to the reported removal difficulty.Video analysis: a twenty-seven (27) second video clip capturing the delivery system post removal was received.The external slider is in the open position with the spiral tube exposed.The user is observed attempting to advance and retract the graft cover by rotating the slider and activating the trigger.The external slider appears to be stuck on the screw gear.The reported difficulties remating the graft cover and taper tip were confirmed on review of the video clip.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information received : it was confirmed the physician had to repair the femoral artery after the difficulty removing the delivery system.It was said the procedure was performed with surgical exposure of the femoral artery so the physician was able to repair without any issue.There was some calcification which created additional steps when the delivery system was stuck.It was clarified that there was no injury to the patient caused by the difficulty removing the delivery system.The surgical exposure of the femoral arteries was already preformed before the device was implanted and these were repaired as planned following delivery system removal.Calcification did not contribute to the difficulty removing the delivery system but there were some calcifications present so it was said it would have been very useful to close the delivery system instead it was removed opened.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis the delivery system returned with the external slider in the home position.The backend wheel was partially opened.There was flattening visible to the graft cover 45.5cm ¿ 49.7cm distal to the strain relief.The graft cover was retracted and advanced using the external slider and the trigger with no issues noted retracting the trigger.The reported removal difficulties could not be confirmed through analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|