THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE RENT; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number L201-90411 |
Device Problems
Display or Visual Feedback Problem (1184); Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was requested, information not yet available.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
|
|
Event Description
|
It was reported that the centrimag patient was getting a computed tomography (ct) scan and after moving back onto their bed, the flow on the centrimag console was not reading a numerical flow, just three dashes.The circuit was examined for kinks, but no obvious kinks were found.Blood flow was confirmed by shining a light on the tubing and checking the motor/pump head for blood flow movement, which confirmed that there was blood flow.The flow sensor was verified to be positioned properly and fully closed.Flow probe connection to the console was verified to be tightly inserted in the centrimag.Despite these steps, no numerical flow reading was achieved.The patient's mean arterial pressure (map) had decreased from 66 to 58 and was holding steady at ~50mmhg, but they had also received a propofol bolus for moving on the ct table, and it was not decreasing.An additional specialist evaluated the situation and was concerned about a flow probe malfunction.The pump head then intermittently made a loud vibratory noise.It was thought that possibly it wasn't seated properly, so the specialists verified that the pump head was "seated" correctly in the motor.The noise from the pump head started to become more frequent but the patient still had stable vital signs.The patient was quickly returned to the cardiac care unit as the pump head began visibly vibrating.The flow, as measured from an ancillary extracorporeal membrane oxygenation (ecmo) device, was significantly reduced.Another flow probe from a non-abbott device was placed onto the centrimag circuit to confirm flow.The decision was made to change the circuit and console/motor system.The changeout was uneventful.The console and pump head were secure for return to abbott.Several hours later in the morning, it was noted that the pump head had a large clot surrounding ~1/4 to 1/3 of the actual pump head "vanes".Related mfr: 2916596-2024-00241.Related mfr: 2916596-2024-00242.Related mfr: 2916596-2024-00652.
|
|
Manufacturer Narrative
|
Correction: the initial report was incorrectly submitted with a cfn-based manufacturer report (mfr) number: 2916596-2024-00240.The mfr number should have been fei-based with the 10-digit fei being 3003306248.Manufacturer's investigation conclusion: the reported event of the centrimag console screen showing a blank flow was confirmed via analysis of downloaded log file; however, the issue was not reproduced during testing of the returned console.The log file contained approximately 7 days of data ( on (b)(6) 2023 ¿ (b)(6) 2024 per the time stamp).On (b)(6) 2024 at 23:21, an s3: system fault alarm activated, correlating to an ifd shutdown sub-fault.After this event, the speed was observed to have dropped to approximately 3700 revolutions per minute and the flow reading was observed to be blank, reading as 0.0 liters per minute.The motor regained its set speed within the same minute; however, the flow remained blank for the remainder of the log file.An f2: flow signal interrupted alarm was observed on (b)(6)2024 at 23:22 and an m4: motor alarm was observed on (b)(6) 2024 at 23:40.These alarms were able to be muted and cleared within a minute of activating; however, the alarms continued to activate until (b)(6) 2024 at 23:44 and (b)(6) 2024 at 23:45.The system was then observed to be manually shut down on (b)(6) 2024 at 23:57 and was removed from patient use for the remainder of the log file.The returned centrimag console (serial number: (b)(6) was evaluated by service depot alongside the returned centrimag flow probe (serial number: (b)(6) and known working test centrimag equipment for several days without any issues or atypical alarms produced.The unit was disassembled to inspect the internal components and no issues were observed.A full functional checkout was performed, and the unit passed all steps of the test without any issues.The serviced and tested centrimag console was returned to the customer site after passing all tests per procedure.The root cause of the reported event could not be conclusively determined through this analysis.The device history records were reviewed, and the records revealed that the centrimag console, serial number (b)(6) , was manufactured in accordance with manufacturing and quality assurance specifications.The 2nd generation centrimag system operating manual (rev.M) section 10 ¿ ¿emergencies/troubleshooting¿ provides instructions for operation when there is a need to exchange the main console or motor with a backup console or motor.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.The 2nd generation centrimag system operating manual (rev.M) section 12.1 ¿ "appendix i ¿ primary console alarms and alerts" cover all alarms (auditory and visual), including system, motor, and flow alarms, and the appropriate actions to take if the issue does not resolve.No further information was provided.The manufacturer is closing the file on this event.
|
|
Search Alerts/Recalls
|
|
|