Catalog Number CDS0706-NTW |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that during device prep of a mitraclip, one of the grippers could not be lowered.Troubleshooting was performed and the issue continued.The clip delivery system (cds) was replaced and the replacement clip was successfully implanted.There was no patient involvement with the issue device.
|
|
Manufacturer Narrative
|
Returned device analysis did not confirm the reported single gripper actuation issue (difficult to open or close).A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.A review of the complaint history did not indicate a lot-specific product issue.Based on information provided and the results of the returned device analysis (unable to confirm the reported issue), a cause for the reported difficult to open or close associated with a single gripper actuation issue during device preparation could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
|
|