• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC QDOT-MICRO, BI-DIRECTIONAL, D-F CURVE, C3, SPLIT HANDLE; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC QDOT-MICRO, BI-DIRECTIONAL, D-F CURVE, C3, SPLIT HANDLE; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Catalog Number D139505
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Fistula (1862)
Event Date 01/05/2024
Event Type  Death  
Manufacturer Narrative
Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturer record evaluation cannot be conducted because the no lot number was provided by the customer.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4) biosense webster manufacturer's reference number (b)(4) has 2 reports.
 
Event Description
It was reported that a patient underwent an atrial fibrillation ablation with a ngen generator and a qdot-micro, bi-directional, d-f curve, c3, split handle and the patient experienced post-op esophageal fistula that resulted in death.The complication was declared several days/weeks after the intervention.The physician's opinion was that use of the qdot-micro contributed to the death but did not elaborate on the issue.The generator was in temperature control mode with a temperature cut off of 55°c.The correct catheter settings were selected on the generator.No error messages observed on biosense webster equipment during the procedure.When asked what modalities were used to prevent esophageal injury, the physician responded with: "temperature in the esophagus".The physician also stated that no perforations occurred during the procedure.This report is for the qdot catheter.Another report has been submitted for the ngen rf generator.
 
Manufacturer Narrative
On 31-jan-2024, bwi received additional information indicating that the previously reported event date of 8-jan-2024 is actually 5-jan-2024.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT-MICRO, BI-DIRECTIONAL, D-F CURVE, C3, SPLIT HANDLE
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18599345
MDR Text Key333990733
Report Number2029046-2024-00322
Device Sequence Number1
Product Code DRF
UDI-Device Identifier10846835016758
UDI-Public10846835016758
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P210027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberD139505
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/08/2024
Initial Date FDA Received01/29/2024
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/21/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTO 3 SYSTEM WITH VISITAG.; NGEN PUMP, EU CONFIGURATION.
Patient Outcome(s) Death; Life Threatening;
-
-