Continuation of d10: w4tr01 crt-p.Implanted (b)(6) 2024.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during the implant procedure, after the left ventricular (lv) lead was placed, it was noted that all vectors that contained the lv1 electrode had high undefined impedance.The physician double checked the connection to the device header, made sure the lead pin was clean and checked through the analyzer however the issue persisted.The physician decided to leave the lead in place and program around the lv1 electrode.All lv impedance alerts were also programmed off.The lv lead was implanted and remains in use.No patient complications have been reported as a result of this event.
|