ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number PM2240 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/09/2024 |
Event Type
malfunction
|
Event Description
|
Related manufacturer report #: 2017865-2024-01931; 2017865-2024-01933 it was reported that the patient was asymptomatic to inhibited paced beats but was dependent on the pacemaker for normal function.The patient was brought into clinic (b)(6) 2024 for a right ventricular (rv) lead revision due to observed dropped pacing beats as a result of noise and oversensing, however the subclavian vein was occluded, and the patient was set for extraction (b)(6) 2024.During the extraction procedure, the right atrial (ra) lead could not be unscrewed from the generator.The ra lead was therefore cut and explanted.The rv lead was noted by the physician to be fractured during the extraction procedure.The generator was replaced due to being close to the normal elective replacement indicator (eri).The patient was stable before, throughout, and post procedure.
|
|
Manufacturer Narrative
|
The reported event of cannot removed the atrial lead from the header was confirmed.Analysis revealed this device has the original scalable brady pacemaker (sbp) header design, which is known to have lead insertion and removal difficulties.The header design is what caused the atrial lead to become stuck and un-removeable from the header.Also, the ventricular lead was pulled apart by the physician while trying to remove it out of the connector.This shows the high abnormal force needed to remove leads from the connectors of the original sbp header design.Actions have already been taken to prevent reoccurrence.A header re-design has been implemented to correct this issue.
|
|
Manufacturer Narrative
|
Correction: section h6: health effect - impact code corrected to "4632 - prolonged surgery".
|
|
Event Description
|
Information received reconfirms that though the pacemaker had reached the elective replacement indicator (eri), there was a suspected issue with the pacemaker set screw such that the right atrial lead could not be unscrewed from the generator.
|
|
Search Alerts/Recalls
|
|
|