Model Number ACT100
Device Problems Device Alarm System (1012); Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2024
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: the reported the reported issues of the instrument failing liquid quality control (qc) and displaying error code 69 were verified during service.The service technician observed error codes 11 and 69 in the statistics log.The issues were resolved by replacing the assy flag sensor.Post-repair testing was performed per specifications.The instrument was serviced/analyzed in the facility by a field service technician.The instrument did not return to a medtronic facility for service/analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that prior to use of a act plus instrument, it was reported that the instrument failed liquid quality control (qc) and displayed error code 69.The instrument was used.There were no adverse patient effects.
 
Manufacturer Narrative
Medtronic received additional information that the customer obtained control values.The customer stated they were not using this instrument in a case at the time.The instrument was replaced.There was no patient involvement, so no adverse effect occurred.Medtronic received additional information that quality control is performed twice a day for the instrument.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACT PLUS INSTRUMENT
Type of Device
TIMER, CLOT, AUTOMATED
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18605737
MDR Text Key334643566
Report Number2184009-2024-00046
Device Sequence Number1
Product Code GKN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K940426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACT100
Device Catalogue NumberACT100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received01/30/2024
Supplement Dates Manufacturer Received02/23/2024
Supplement Dates FDA Received03/20/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-