A philips remote service engineer (rse) spoke by telephone support with the customer biomed who performed an audio test on the device, which did not produce any sound.Based on the information available and the testing conducted, the rse confirmed the cause ofthe reported problem was the speaker.The reported problem was confirmed.The rse instructed the biomed to clear the device data and reboot the device.Additionally, the rse advised the biomed to replace the mainboard and speaker assembly if the inop remains.The investigation concludes that no further action is required at this time.
|