• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2 Back to Search Results
Model Number 863283
Device Problems No Audible Alarm (1019); No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2024
Event Type  malfunction  
Event Description
The customer reported a speaker malfunction inop message and the device was unable to produce sound.The device was not in clinical use at the time of the event.No adverse event or patient harm was reported.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information and the complaint is still under investigation.A follow-up report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
A philips remote service engineer (rse) spoke by telephone support with the customer biomed who performed an audio test on the device, which did not produce any sound.Based on the information available and the testing conducted, the rse confirmed the cause ofthe reported problem was the speaker.The reported problem was confirmed.The rse instructed the biomed to clear the device data and reboot the device.Additionally, the rse advised the biomed to replace the mainboard and speaker assembly if the inop remains.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESIGNS VS4 NBP, SPO2
Type of Device
SURESIGNS VS4 NBP, SPO2
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18606655
MDR Text Key335081595
Report Number1218950-2024-00061
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838087095
UDI-Public00884838087095
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number863283
Device Catalogue Number863283
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2024
Initial Date FDA Received01/30/2024
Supplement Dates Manufacturer Received02/06/2024
Supplement Dates FDA Received03/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-