Model Number SE3690
Device Problems Failure to Calibrate (2440); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Event Description
According to the reporter, current leakage exceeds over 2ma.There was no patient involvement.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
New information has been received, and reassessment of the complaint found that it is no longer a reportable event.The event is no longer associated with a serious injury or potential for serious injury with reoccurrence.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found a functional evaluation found that the motor fails hi pot.It was reported that the device was out of specification.The reported issue was confirmed.The most likely cause was traced to component failure.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDVAC
Type of Device
APPARATUS, EXHAUST, SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18607186
MDR Text Key334616885
Report Number1717344-2024-00292
Device Sequence Number1
Product Code FYD
UDI-Device Identifier10884524001555
UDI-Public10884524001555
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K980915
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSE3690
Device Catalogue NumberSE3690
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/08/2024
Initial Date FDA Received01/30/2024
Supplement Dates Manufacturer Received03/29/2024
Not provided
Supplement Dates FDA Received03/29/2024
05/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-