• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. 1.5T LINX, 14B; ANTI-REFLUX IMPLANT Back to Search Results
Catalog Number LXMC14
Device Problems Device Appears to Trigger Rejection (1524); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Dysphagia/ Odynophagia (1815); Vomiting (2144)
Event Date 01/18/2024
Event Type  Injury  
Manufacturer Narrative
(b)(4).Date sent: 1/30/2024.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Lot number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.Additional information was requested, and the following was obtained: prior to linx placement, did the patient have an egd, ph, and manometry studies done? if yes, could you please share the results? yes, please see attached.On what date did the implant take place? (b)(6) 2021.On what date did the explant take place? (b)(6) 2024.What is the lot number of the linx device? 26414.When using the linx sizing device what technique was used to determine the size? 3 measurements.Did the patient have an autoimmune disease? no.Is the patient currently taking steroids / immunosuppressive drugs? no.Did the patient have any pre-existing dysphagia or other conditions (other than gerd)? hiatal hernia.How severe was the dysphagia/odynophagia before intervention? were there any intra-operative complications during implant? no.Was there any hiatal or crural repair done at the same time as the implant? yes, hiatal hernia repair.Were there any other contributing factors that led to the removal of the device other than the reported dysphagia? food getting stuck and vomiting.Patient had two endoscopies w/dilation and had no significant change with his symptoms.Besides the reported dysphagia, what was the reason for removal of the linx device? food getting stuck and vomiting.Was the device found in the correct position/geometry at the time of removal? yes.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported a linx removal for mild dysphagia.It¿s been in for over two years.Normal anatomy.No real improvement with dilation.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
(b)(4).Investigation summary: overall review of the device function and dimensions show no anomalies from a device that has been reasonably changed as part of the explant procedure.Visual analysis was consistent with an explanted device, and link length and tensile force were found to meet the applicable specifications.Overall, no analysis conclusions relevant to the patient experience were found.
 
Manufacturer Narrative
(b)(6) date sent: 3/5/2024 photo analysis: endoscopy images were reviewed by a medical safety officer.As per medical safety officer: "the endoscopy images demonstrate evidence of impaired esophageal dysmotility and dilation." the mechanism/cause of failure is unknown as a hands-on analysis of the device is necessary to determine the cause of failure.No further investigation can be completed at this point.A manufacturing record evaluation was performed for the finished device 26414 number, and no non-conformances related to the malfunction were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5T LINX, 14B
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4545 creek rd.
cincinnati, OH 45242
3035526892
MDR Report Key18608214
MDR Text Key334141353
Report Number3008766073-2024-00023
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005349
UDI-Public00855106005349
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/11/2023
Device Catalogue NumberLXMC14
Device Lot Number26414
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/30/2024
Supplement Dates Manufacturer Received02/29/2024
03/07/2024
03/25/2024
Supplement Dates FDA Received03/05/2024
03/08/2024
03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-