Catalog Number 11011-29
Device Problems Difficult to Advance (2920); Device Dislodged or Dislocated (2923); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that prior to the procedure during insertion of the 8.0mmx29mmx135cm omnilink elite stent delivering system (sds) into the introducer sheath, the stent dislodged but remained on the balloon and was noted to have flared struts.The sds was prepped after the sheath/stylet removal with no force or manipulation noted during removal.Another abbott device was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was returned for analysis.The reported stent dislodgement and material deformation were confirmed.The reported difficulty to advance could not be confirmed due to the condition the device was received for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the device interacted with the introducer sheath during advancement causing the reported difficulty to advance and subsequent stent dislodgement with material deformation (misaligned, bent, stretched and overlapping stent struts).There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
Subsequent to the initially filed mdr report, it was reported that resistance was felt during advancement into the introducer sheath.The stent delivery system was removed from the introducer sheath and the stent dislodged outside the anatomy.Additionally, the stent was noted to have flared struts.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Type of Device
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18610634
MDR Text Key334143225
Report Number2024168-2024-01230
Device Sequence Number1
Product Code NIO
UDI-Device Identifier08717648142727
UDI-Public(01)08717648142727(17)260331(10)2100341
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P110043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number11011-29
Device Lot Number2100341
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/08/2024
Initial Date FDA Received01/30/2024
Supplement Dates Manufacturer Received03/26/2024
Supplement Dates FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-