|
Model Number 45026 |
Device Problems
Entrapment of Device (1212); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/04/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the catheter stuck, and tip frayed occurred.An angiojet avx catheter was advanced for a thrombectomy procedure.During the procedure, the physician noticed that the radiopaque markers were suddenly farther apart than normal.The device was pulled out to investigate, but the catheter was stuck and would not come back out of the sheath.It was stuck to the.035 non-boston scientific glidewire and could not get either of the device back out.The three devices had to be removed all together out of the patient at the same time.Once the physician had pulled the wire, the catheter and the sheath out together, a portion of the wire was still in the patient.They were able to free the catheter and sheath from the wire and inserted a new sheath to maintain access and was able to remove the wire.Moreover, it was noted that the catheter tip was frayed and mangled.The procedure was completed this device as they are far along in the case and had gotten enough clot out and did not need another catheter or product.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: returned product consisted of an angiojet avx catheter.Media was returned in the form of a photo.The photo was reviewed which showed the tip damage which was confirmed during analysis.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed shaft kinks at 42 cm and 37.5 from proximal marker band.No guidewire was received with the device and no guidewire was inserted due to the extreme damage to the device.Functional testing was not attempted due to the extreme damage to the device.During analysis, damage was observed measuring from the proximal marker band at 1.5 cm to 4.5 cm, the entire section of the shaft was stretched with the hypotube detached and separated at 1.5 cm from the proximal marker band.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Search Alerts/Recalls
|
|
|