COOK INC; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problems Unraveled Material (1664); Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Event Description
It was reported that the wire guide from a cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set kinked.When the wire guide was advanced into the patient, resistance was felt.The wire was removed, and kinking was noted.A new device set was obtained to complete the procedure.It was noted the patient did not have any tortuous anatomy.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Upon receipt of the complaint device, unravelling of the wire guide was observed, thus prompting this report.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.D1 - brand name: cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter set.This report includes information known at this time.A follow-up report will be submitted should additional, relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18615043
MDR Text Key335096710
Report Number1820334-2024-00140
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00827002478299
UDI-Public(01)00827002478299(17)250412(10)14830109
Combination Product (y/n)Y
PMA/PMN Number
K081113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberC-UTLM-701J-RSC-ABRM-HC-RD
Device Lot Number14830109
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/29/2024
Initial Date FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexFemale
Patient Weight60 KG
-
-