• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 802.11A/B/G

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 802.11A/B/G Back to Search Results
Model Number 865352
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Event Description
The customer reported that the system speaker is defective.The device was in use on patient at time of event, there was no adverse event reported.
 
Manufacturer Narrative
A philips field service engineer (fse) went onsite to evaluate the device in question.The fse indicated during an evaluation of the system that the displays a telemetry error message.The philips field service engineer (fse) replaced the telemetry transmitter and tested the settings.The device was operational after repairs were completed and the device was returned to the customer.The investigation concludes that no further action is required at this time.(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX40 802.11A/B/G
Type of Device
INTELLIVUE MX40 802.11A/B/G
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18615249
MDR Text Key334286225
Report Number1218950-2024-00062
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838030350
UDI-Public00884838030350
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865352
Device Catalogue Number865352
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/04/2024
Initial Date FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-