This report is being supplemented to provide additional information based on the legal manufacturer's investigation and correction to h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the disconnected biopsy port from the scope was it was not attached between control body assembly and proximal port during the bedside cleaning due to stress.The event can be detected/prevented by following the instructions for use which state: ¿instructions: oes cystonephrofiberscope, olympus cyf-5, olympus cyf-5a.Important information ¿ please read before use, [warning]: do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, or control section.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, or light guide cable with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.¿.Olympus will continue to monitor field performance for this device.
|