Model Number RFIT-ASY-0147
Device Problem False Positive Result (1227)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2023
Event Type  malfunction  
Event Description
Summary: please note that the customer used an expired biofire bcid2 panel pouch and is considered an off-label use.I.M.Alliance sarl- clinical (morocco) reported a potential false positive candida tropicalis result on the biofire blood culture identification 2 (bcid2) panel after testing a patient blood culture sample.Due to the biofire bcid2 panel result, it is unknown if the patient was harmed or affected.The investigation determined that the most likely cause of the discrepant c.Tropicalis result was the presence of non-viable organism/nucleic acid in the blood culture media bottle.
 
Manufacturer Narrative
Investigation: please note that the customer used an expired biofire bcid2 panel pouch and is considered an off-label use.A potential false positive c.Tropicalis result on the biofire bcid2 panel was reported after testing a patient blood culture sample on (b)(6) 2023.The patient impact due to the biofire bcid2 result is unknown.No serious injury or death occurred.Conclusion: the investigation determined that the most likely cause of the discrepant c.Tropicalis result was the presence of non-viable organism/nucleic acid in the blood culture media bottle.Similar events were recently observed in the field with the combination of the biofire bcid2 panel and bd bactec¿ blood culture bottles.A field safety corrective action (fsca) was issued in the u.S.On january 26, 2024, and countries outside u.S.On january 30, 2024, via fsca (b)(4) for this issue.While blood culture vials are autoclaved and routinely quality controlled for sterility, non-viable organisms or nucleic acids can remain in the blood culture media after the sterilization process.The presence of non-viable organisms and nucleic acid does not compromise the intended use of blood culture media, culturing viable microorganisms; however, the biofire bcid2 panel does not distinguish between nucleic acid from viable or non-viable organisms.The biofire bcid2 panel is working as intended.The "laboratory precaution" and "limitation" sections of the biofire bcid2 panel instructions for use (ifu) (www.Online-ifu.Com/iti0048) outlines the potential for false positive detections during molecular testing with sterile blood culture media containing detectable levels of non-viable organisms and/or nucleic acid.Importantly, results from the biofire bcid2 panel are intended to be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.All identification results provided by the biofire bcid2 panel are intended to be interpreted in conjunction with gram stain results.Clinical performance: according to table 38.Biofire bcid2 panel clinical performance summary, candida spp.Of the biofire bcid2 panel ifu, the performance claim for the c.Tropicalis assay compared to standard of care identification for genus level followed by pcr & sequencing of isolates for species identification showed an overall sensitivity of 100% (95% ci 93.5-100%) and an overall specificity of 99.9% (95% ci 99.7-100%).The single false positive specimen was identified as a cross-reactivity between the biofire bcid2 panel c.Tropicalis assay and high titer candida parapsilosis.This cross-reactivity is a known limitation of the biofire bcid2 panel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL
Type of Device
BIOFIRE® BLOOD CULTURE IDENTIFICATION 2 (BCID2) PANEL
Manufacturer (Section D)
BIOFIRE DIAGNOSTICS, LLC
515 colorow dr
salt lake city UT 84108
Manufacturer (Section G)
BIOFIRE DIAGNOSTICS, LLC
515 colorow dr
salt lake city UT 84108
Manufacturer Contact
515 colorow dr
salt lake city, UT 84108
8017366354
MDR Report Key18619389
MDR Text Key334882402
Report Number3002773840-2024-00212
Device Sequence Number1
Product Code PAM
UDI-Device Identifier00815381020338
UDI-Public00815381020338
Combination Product (y/n)N
Reporter Country CodeMO
PMA/PMN Number
K193519
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/20/2023
Device Model NumberRFIT-ASY-0147
Device Catalogue NumberRFIT-ASY-0147
Device Lot Number1420022
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/02/2024
Initial Date FDA Received01/31/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-