A distributor reported on behalf of a healthcare facility in china, via a fisher & paykel (f&p) healthcare representative, that three (b)(6) vented autofeed humidification chambers provided with rt265 infant dual-heated evaqua2 breathing circuits were found leaking water prior to patient use.There was no patient involvement.
|
Ps440024
corrections:
section d: updated d1, d2, d4 to state mr290v vented autofeed humidification chamber as the suspect medical device.Section g: updated 510(k) to k934140
section d4 & h4:
device 1-2: lot#: 2102698415 ; date of manufacturing; 19 july 2023; udi: (01)09420012430205(10)2102698415(11)230719
device 3: lot#: 2102667465; date of manufacturing; 26 june 2023; udi: (01)09420012430205(10)2102667465(11)230626
section h11:
method: the subject mr290v vented autofeed humidification chambers were not returned to fisher & paykel healthcare for evaluation, therefore, no investigation could be performed.Conclusion: without the subject devices, we are unable to determine what may have caused the reported leakage.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290 chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following:
- "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers."
- "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient."
- "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient."
- "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
|
A distributor reported on behalf of a healthcare facility in china, via a fisher & paykel (f&p) healthcare representative, that three mr290v vented autofeed humidification chambers provided with rt265 infant dual-heated evaqua2 breathing circuits were found leaking water prior to patient use.There was no patient involvement.
|