LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2024
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in smithfield london, united kingdom.A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue.Pump was ran clockwise and anti-clockwise with over occlusion and under occlusion, with no tubing inserted and tubing inserted and the reported fault was not reproduced.Further functional verification testing was completed without issues and the unit was returned to service.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2013.Based on all information collected, it cannot be ruled out that the most likely root cause of reported event was an intermittent communication failure between the computer board (hkr) and the shaft angle encoder.
 
Event Description
Livanova deutschland received a report that a s5 roller pump gave an error message related to shaft angle encoder fault during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18620903
MDR Text Key335409889
Report Number9611109-2024-00051
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 01/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/03/2024
Initial Date FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-