• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had a pressure sensor failure.There was no patient involvement when the issue occurred.No patient or user harm reported.A field service engineer (fse) evaluated the device, and reported that the event log was reviewed, and noticed seven occurrences of error code 110b ((cbit) check vent: proximal pressure sensor auto zero failed).The fse then removed the gas delivery system (gds) from the device, and notice that six of the solenoid pins were not seated into the data acquisition (da) board; it was also noted that the pins were bent.The fse reported that they replaced all solenoid valves, and also ran a performance verification test (pvt).The device passed all testing and returned to the customer for full use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18622243
MDR Text Key334318044
Report Number2518422-2024-05436
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received02/01/2024
Date Device Manufactured10/21/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-