Model Number 799-EMVP-05-01 |
Device Problems
Mechanical Problem (1384); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during training by facility staff, the device displayed "compressor fault- 2001", "internal comm failure - 1175, "internal comm failure- 1472", and "internal comm failure -1474" error messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the customer's report was observed during review of the device history log.The device was put through extensive testing including stress testing, trouble shooting, and an internal inspection without duplicating the report.The central processing unit (cpu) to power interface module (pim) board ribbon cable was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|