Model Number CLV-190
Device Problems Arcing (2583); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the light source exhibited internal arcing and clicking sound when it came on, electrical hazard.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned to olympus facility for evaluation.In addition, evaluation found the following: the caution label was worn; front panel overlay was peeling; the device had a faulty lamp igniter; and the scope socket was bad due to a loose locking mechanism.The investigation is ongoing.A supplemental report will be submitted if any additional information is provided by the user facility.
 
Manufacturer Narrative
This supplemental report is being submitted to correct information that was provided in the initial medwatch report and to provide the results of the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, the root cause for the event could not be determined.However, it is likely that abnormal noise occurred when the power of the lamp was turned on due to faulty ignitor.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18625792
MDR Text Key334777976
Report Number3002808148-2024-01017
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/30/2024
Initial Date FDA Received02/01/2024
Supplement Dates Manufacturer Received02/14/2024
Supplement Dates FDA Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-