• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRECISION SPINE, INC REFORM PEDICLE SCREW SYSTEM; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRECISION SPINE, INC REFORM PEDICLE SCREW SYSTEM; SCREWDRIVER Back to Search Results
Catalog Number 39-RD-0060
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Manufacturer Narrative
H3 device evaluation - the driver was returned with a fracture of the tip.The fracture plane is skewed relative to the driver's longitudinal axis.This is indicative of parts failing under combination loading.It is believed that bending moments were applied in combination with torque which led to this failure.The part has likely seen extensive use since being fabricated in 2014 so this failure may have been initiated during prior usage.It is unclear if a counter torque wrench was being used during set screw tightening.Proper use of a counter torque wrench helps mitigate bending moments acting on the driver's tip.Review of device history records found ten (10) pieces of lot 0127it released for distribution on (b)(6) 2014 with no deviation or anomalies.Two-year complaint history review did not reveal a trend for reports of this nature for the reported part number.No corrective actions are recommended at this time.
 
Event Description
It was reported that a procedure was performed on (b)(6) 2023, utilizing the reform pedicle screw system.While final tightening the last lock screw the torque limiting handle torqued out after the initial two clicks, so the surgeon gave it some extra turns resulting in the tip of the t25, lock-screw torque driver, reform (39-rd-0060) broke.The broken piece was recovered, and the case was completed utilizing other instrumentation readily available in the set.There was no patient injury or delay to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFORM PEDICLE SCREW SYSTEM
Type of Device
SCREWDRIVER
Manufacturer (Section D)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer (Section G)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer Contact
mike dawson
2050 executive drive
pearl, MS 39208
6014204244
MDR Report Key18625851
MDR Text Key334347443
Report Number3005739886-2024-00007
Device Sequence Number1
Product Code HXX
UDI-Device Identifier00840019928366
UDI-Public00840019928366
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number39-RD-0060
Device Lot Number0127IT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2024
Initial Date FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-