It was reported this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4 and flail.An xt clip was inserted and grasping was performed.However, the clip became caught in chordae and a gripper actuation issue was observed.Troubleshooting was performed and the clip was removed from chordae.The physician then decided to remove the clip and replace it.Upon removal, it was observed tissue damage had occurred.An additional xtw clip was then inserted and implanted without issues.However, mr did not reduce and remained at a grade of 4.No clinically significant delay in the procedure.
|
All available information was investigated, and the reported single gripper actuation issue was confirmed via returned device analysis.Additionally, the gripper line was observed to be broken.The reported difficult to remove from anatomy could not be replicated in a testing environment.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and the returned device analysis, the cause of the reported difficult to remove from anatomy was unable to be determined.The observed broken gripper line was likely a result of clip being caught in anatomy and troubleshooting.The reported single gripper actuation issue was a cascading event of the reported difficult to remove from anatomy.The reported tissue injury was a cascading event of the reported difficult to remove from anatomy.The reported patient effect of tissue injury, as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.The reported serious injury/ illness/ impairment was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|