HOLOGIC, INC BREVERA SYSTEM; BIOPSY SYSTEM Back to Search Results
Model Number BREV100
Device Problem Failure to Obtain Sample (2533)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  Injury  
Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.H3: the device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event Description
It was reported that on (b)(6) a brevera procedure was performed and an issue with the suction occurred and the system was not retrieving samples.The doctor could insert and fire the needle, but no cores were cut, and the procedure was rescheduled.A field engineer examined the equipment and was not able to reproduce the error, the field engineer replaced the filter and ran testing.The device worked as per the manufacturer's specifications.No additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREVERA SYSTEM
Type of Device
BIOPSY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
ariel lafuente
562 parkway
coyol free zone building b24
san jose 20102
CR   20102
MDR Report Key18627502
MDR Text Key334368431
Report Number1222780-2024-00034
Device Sequence Number1
Product Code KNW
UDI-Device Identifier154200455075
UDI-Public(01)154200455075(17)11220209
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBREV100
Device Catalogue NumberBREV100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2024
Initial Date FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-