• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40 SYSTEM SILVER; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40 SYSTEM SILVER; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number R1111177
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  malfunction  
Event Description
Loss of power alarm.A ventilator was returned to the manufacturer for routine preventative maintenance.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, a failure of the "loss of power" alarm to sound was observed.The device's main board was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40 SYSTEM SILVER
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18628461
MDR Text Key334375313
Report Number2518422-2024-05534
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121623
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR1111177
Device Catalogue NumberR1111177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2024
Initial Date Manufacturer Received 01/25/2024
Initial Date FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-