CHRISTOPH MIETHKE GMBH & CO KG M.BLUE LUMBAL 5; HYDROCEPHALUS MANAGEMENT Back to Search Results
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964)
Patient Problem Hydrocephalus (3272)
Event Date 12/19/2023
Event Type  Injury  
Event Description
It was reported that a m.Blue (#fx851t) was implanted during a procedure performed on (b)(6) 2023.According to the complainant, the valve showed an under-drainage, slowed outflow ex vivo.The patient had another valve implanted during the operation on (b)(6) 2023.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 68 years.Weight: 90 kilograms.Height: 165 centimeters.Gender: female.
 
Manufacturer Narrative
Visual inspection: during the investigation, no significant deformations or damage of the valve was determined.Permeability test: a permeability test has shown that the m.Blue is permeable and that the lumbalcatheter has a blockage.The catheter was tested for permeability in sections and organic particles were flushed out.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the m.Blue operates within the accepted tolerance in the horizontal position, but not in the vertical position.An accelerated outflow of m.Blue could be determined.Adjustment test: the m.Blue valve was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Results: based on our examination results, we can detect organic particles in the catheter.The particles may have led to the malfunction complained about, and the m.Blue works as expected.Deposits caused by natural substances in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of proteins can impair the integrity of the valve.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE LUMBAL 5
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18629172
MDR Text Key334408923
Report Number3004721439-2024-00005
Device Sequence Number1
Product Code JXG
UDI-Device Identifier04041906506094
UDI-Public4041906506094
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K192266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/08/2024
Initial Date FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexFemale
Patient Weight90 KG
-
-