Model Number 16-02-85 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2024 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland has received a report that, a 3t heater cooler displayed error pump 3 is blocked (e22) on patient side during a procedure.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united states.Customer reported the patient side indicated an error e22.Confirmed system patient pump 2 will not turn when activated.Replaced the entire patient pump bridge.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
A complaint database review was performed and identified that unit was manufactured in 2014 and no other similar event was reported since its installation.Based on technical intervention and on investigation results of similar events, the potential root causes of the pump block could be: (i) damage of the motor ball bearing due to corrosion caused by the environment in which the pump is located (high temperatures and high level of humidity) and / or (ii) bush abrasion due to shaft misalignment.
|
|
Event Description
|
See initial report.
|
|