It was reported that a percutaneous nephrolithotomy was performed on (b)(6) 2024.After opening the package of balloon dilation catheter, the doctor found that the gauge of the pressure pump was disconnected and fell off, so they replaced it with a new product and performed the operation.
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
The reported event is confirmed cause unknown.Visual evaluation noted one inflation device was returned to atrion inside a zip lock bag with the inflator tray.Upon receipt of the complaint device a visual inspection was conducted.During the visual inspection, it was noted that the gauge was broken from the inflation device housings at the glue plug area.The needle on the gauge was also noted to be within the zero position.Also received 1 photo sample which shows top overview of eagle inflation device.Based on physical sample received the product does not meet specifications.A potential root cause for this type of failure could be inappropriate package design.However, there was insufficient information to confirm this potential root cause.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "instructions for use caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Description the x-force® nephrostomy balloon dilation catheter is a dual lumen catheter with a 24 (8mm) or 30fr (10mm) balloon mounted on the distal tip.It has a radiopaque tip and a radiopaque marker beneath the balloon.The lumen labeled with rated burst pressure (xx atm) is for balloon inflation.The other lumen allows the catheter to track over a 0.038¿ (.97mm) diameter guidewire and can be used for monitoring of pressure or the infusion of medication and/or contrast medium.Each balloon inflates to a stated diameter and length at a specific pressure ¿ typically at 10 atm.The balloon dilation catheter comes packaged with a refolding tool and a working sheath.It is available with or without an inflation device.It comes sterile and is for single use only.Indications for use the x-force® nephrostomy balloon dilation catheter is recommended for use in the dilation of the nephrostomy tract and for placement of the working sheath.Contraindications do not use the x-force® nephrostomy balloon dilation catheter in the presence of conditions which create unacceptable risk during the dilation of the nephrostomy tract.Warnings: ¿ if resistance is felt when removing either the catheter or the guidewire from the working sheath, stop and consider removing them as a single unit to prevent damage to the product.Applying excessive force to the catheter can result in tip breakage or balloon separation.¿ do not use air or any gaseous substances as a balloon inflation media, always use sterile liquid media.¿ this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: ¿ only a physician who has an understanding of the clinical applications, technical principles and associated risks associated with balloon dilation of the nephrostomy tract should use this device.¿ after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable laws and regulations." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
It was reported that a percutaneous nephrolithotomy was performed on (b)(6) 2024.After opening the package of balloon dilation catheter, the doctor found that the gauge of the pressure pump was disconnected and fell off, so they replaced it with a new product and performed the operation.
|