• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Model Number ASKU
Device Problems Crack (1135); Device Damaged by Another Device (2915)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that during an intraocular lens (iol) implant procedure, it was noted that three lenses as they came through the cartridge, the cartridge cracked and seemed to have scratched the lenses.The lenses were explanted during initial procedure.Additional information has been requested.
 
Manufacturer Narrative
The product was not returned.Complaint history and product history records could not be reviewed because the reporter did not provide a lot number or any identification traceable to the manufacturing documentation.Associated products were not provided.It is unknown if qualified associated products were used.The product investigation could not identify a root cause for the reported complaint.The product was not returned.Not enough information was provided for further investigation.The instructions for use (ifu) instructs: the company intraocular lens (iol) delivery system is for implantation of qualified company foldable iols.No unqualified lenses should be used with the company iol delivery system.The company cartridges are qualified for use with compatible company handpieces for the surgical implantation of company qualified foldable iols.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Use the company cartridge at operating room temperatures between 18° c (64° f) and 23° c (73° f).The reporting facility has not responded to requests for follow up.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH IOL DELIVERY SYSTEM, CARTRIDGE
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18631963
MDR Text Key334491637
Report Number1119421-2024-00189
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/08/2024
Initial Date FDA Received02/02/2024
Supplement Dates Manufacturer Received03/14/2024
Supplement Dates FDA Received04/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-