• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG M.BLUE PLUS VALVE; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG M.BLUE PLUS VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX804T
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964)
Patient Problem Hydrocephalus (3272)
Event Date 01/12/2024
Event Type  Injury  
Event Description
It was reported that a m.Blue plus (#fx804t) was implanted during a procedure performed on (b)(6) 2024.According to the complainant, the shunt system showed an under-drainage and adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2024.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 28 years.Weight: 100 kilograms.Height: 166 centimeters.Gender: female.
 
Manufacturer Narrative
Visual inspection: during the investigation, no significant deformations or damage of the valves were determined.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The results show that the valves operate within the permissible tolerance in their respective relevant position.Adjustment test: during the adjustment test it was determined that both valves are adjustable in all pressure ranges.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected; and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, no deposits were found in both valves.To make the deposits in the shunt system more visible, they were colored using a staining solution.Results: based on our investigation results, we cannot determine functional deviations or other abnormalities in the product.How the abovementioned functional impairment occurred is not clear to us at the time of examination.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE PLUS VALVE
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18636126
MDR Text Key334491084
Report Number3004721439-2024-00020
Device Sequence Number1
Product Code JXG
UDI-Device Identifier04041906504793
UDI-Public4041906504793
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K192266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFX804T
Device Catalogue NumberFX804T
Device Lot Number20063301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
Patient SexFemale
Patient Weight100 KG
-
-