Evaluation of the returned catrx confirmed a fracture.Further evaluation revealed that the catheter was kinked at the fractured location, indicating that the device likely kinked prior to fracturing.This damage typically occurs if the device is advanced against resistance.The device may become kinked and may subsequently fracture.Further evaluation also revealed kinks and ovalizations along the length of the device.This damage was incidental to the complaint and likely occurred during removal of the device from the procedure and during packaging for return to penumbra.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|
The patient was undergoing a thrombectomy procedure in the superficial femoral artery (sfa), popliteal artery, and anterior tibial artery using an indigo system catrx aspiration catheter (catrx), a non-penumbra sheath, a non-penumbra microcatheter, and a guidewire.During the procedure, the physician completed one pass using the catrx.The catrx was removed to perform an angiogram.Upon removal, the physician noticed that the catrx was fractured at the distal end.It was reported that the angiogram revealed that the distal end of the catrx migrated to the distal anterior tibial artery.The physician attempted to remove the fractured catrx distal end using a snare device; however, the catrx distal end could not be removed.The thrombectomy procedure had ended at this point.The patient was taken to another hospital and an open surgery was performed to remove the fractured distal end of the catrx.
|