BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
|
Back to Search Results |
|
Model Number DB-2202-45 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a deep brain stimulation (dbs) lead contact was damaged during a permanent implant procedure.The procedure was successfully completed with an alternate device.The returned lead revealed that the distal array is fractured between electrode 1 and electrodes 2,3, and 4.There was no patient harm due to the reported event.
|
|
Manufacturer Narrative
|
The returned lead was analyzed and visual inspection revealed that the distal array is fractured between electrode 1 and electrodes 2,3, and 4.A product labeling review identified that the device was used per the instructions for use (ifu) product label and it did not reveal any anomalies.Based on all available information, the reported event of the lead damage was confirmed.The returned lead was analyzed and visual inspection revealed that the distal array is fractured between several lead electrodes.Therefore, with the reported observations and the labeling review, it has been determined that the probable cause was a manufacturing deficiency.
|
|