• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD INSYTE AUTOGUARD; PERIPHERAL IV CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. BD INSYTE AUTOGUARD; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 38182314
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2023
Event Type  malfunction  
Manufacturer Narrative
A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event Description
It was reported that bd insyte autoguard hub did not connect with adaptor.The following information was provided by the initial reporter, translated from portugese to english: when i performed the venipuncture on the patient, when i connected the micro cleave connector to the jelco n° 22, it didn't screw in (it connected to each other), so i had to put another connector on, which also didn't connect to the jelco.I had to remove the access and perform a new puncture on another limb.
 
Manufacturer Narrative
A device history record review was completed for provided material number 38182314 and lot number 2209115.The review did not reveal any detected abnormalities during the production process that could have contributed to the reported defect and all quality tests were found to be within specification.As neither physical samples nor picture samples were available for return, a thorough sample analysis could not be completed.If the affected samples become available for this incident or any potential future incidents, we would greatly appreciate the opportunity to review them.Based on the investigation results, an exact cause could not be determined for the reported event.At this time, further action has not been determined necessary.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOGUARD
Type of Device
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora
BR 
Manufacturer (Section G)
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
ave. pres.juscelino kubitschek
273 francisco bernardino
juiz de fora 81460
BR   81460
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18637535
MDR Text Key335411924
Report Number9610048-2024-00005
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903818235
UDI-Public(01)00382903818235
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number38182314
Device Lot Number2209115
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2024
Initial Date FDA Received02/04/2024
Supplement Dates Manufacturer Received05/23/2024
Supplement Dates FDA Received05/26/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/08/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-