Catalog Number VMC9694
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a buretrol sol.Admin.Set cable fell apart which resulted in a leak.This occurred prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured in march 2023.H10: the actual device was not available; however, two photographs of the sample and retained samples were provided for evaluation.The sample photographs underwent visual inspection, and it was observed that the spike was disconnected from the tube.The reported condition was verified.The cause of the reported condition could not be determined.Visual inspection of the retention samples did not identify any abnormalities that could have contributed to the reported condition.Functional underwater leak and gravity tests were performed, and no issues were observed.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURETROL SOLUTION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TUNISIA
route de chebbaou
2021oued e
tunis
TS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18640178
MDR Text Key334540113
Report Number1416980-2024-00322
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberVMC9694
Device Lot Number23C17T164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received02/05/2024
Supplement Dates Manufacturer Received03/01/2024
Supplement Dates FDA Received03/07/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-