H4: the lot was manufactured in march 2023.H10: the actual device was not available; however, two photographs of the sample and retained samples were provided for evaluation.The sample photographs underwent visual inspection, and it was observed that the spike was disconnected from the tube.The reported condition was verified.The cause of the reported condition could not be determined.Visual inspection of the retention samples did not identify any abnormalities that could have contributed to the reported condition.Functional underwater leak and gravity tests were performed, and no issues were observed.The reported condition was not verified on the retained samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|