• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS ELITE; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS ELITE; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 10676
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2024
Event Type  malfunction  
Event Description
It was reported that stent dislodgment occurred.The 90% stenosed target lesion was located in the severely tortuous and moderately calcified left anterior descending artery.Following pre-dilatation of a 2.5 x 12 semi compliant balloon catheter, a 32 x 3.00 promus elite drug-eluting stent was advanced for treatment.Due to a very tortuous anatomy of the vessel, the stent failed to track the lesion even after trying multiple times.However, when the physician tried removing the stent, the stent came off of the balloon intact near the hub of a non-boston scientific (bsc) guide catheter.A non-bsc stent was used, and the procedure was complete successfully.No patient complications were reported.
 
Manufacturer Narrative
E1: initial reporter address 1: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELITE
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18640251
MDR Text Key334542738
Report Number2124215-2024-04728
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10676
Device Catalogue Number10676
Device Lot Number0031832091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/20/2024
Initial Date FDA Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-