|
Model Number 45031 |
Device Problems
Break (1069); Device Difficult to Setup or Prepare (1487); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/09/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that catheter break occurred.An angiojet solent omni was used for thrombectomy procedure.During the procedure, the device stopped priming at 10s.The device still could not continue to prime after reset.During inspection, it was noted that there was a rupture in the catheter and the shaft was kinked.The device had not entered the patient when it ruptured, and the procedure was completed with another of the same device.No complications were reported, and the patent was stable post procedure.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the device was returned for evaluation.Media was returned in the form of videos.The videos were reviewed which showed a supply line separation.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed no damage.The supply/effluent line showed a severe kink and broken supply line.Functional testing was attempted; however, the device would not prime.During analysis at the severely kinked location at 29 cm from the hub/manifold, the supply line was broken and completely separated.The device could not be functionally tested due to the extreme damage on the device.No pressure reading was reported before the console would stop the priming process and issue an under-pressure alarm.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that catheter break occurred.An angiojet solent omni was used for thrombectomy procedure.During the procedure, the device stopped priming at 10s.The device still could not continue to prime after reset.During inspection, it was noted that there was a rupture in the catheter and the shaft was kinked.The device had not entered the patient when it ruptured, and the procedure was completed with another of the same device.No complications were reported, and the patent was stable post procedure.
|
|
Search Alerts/Recalls
|
|
|