• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS CORPORATION LLD EZ LEAD LOCKING DEVICE; STYLET, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS CORPORATION LLD EZ LEAD LOCKING DEVICE; STYLET, CATHETER Back to Search Results
Model Number 518-062
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 01/17/2024
Event Type  Injury  
Event Description
A lead extraction procedure commenced to remove a right atrial (ra), a right ventricular (rv) and a left ventricular (lv) lead due to non-function.Spectranetics lld ez lead locking devices (lld ezs) were inserted into each lead to provide traction.Beginning with a spectranetics 14f glidelight laser sheath with a medium visisheath dilator sheath on the ra lead, progress stalled due to lead on lead binding.Moving to the lv lead with the same glidelight, the lead was successfully removed.Moving again to the ra lead using the glidelight and visisheath, the remaining lead on lead binding was resolved in the innominate and superior vena cava (svc) regions.While the glidelight and visisheath were still located in the innominate/svc region, the ra lead came free and was removed, along with the glidelight and visisheath; however, the patient''s blood pressure dropped.Rescue efforts began, including rescue balloon and sternotomy.A right atrial appendage (raa) perforation was discovered, clamped, and repaired.Once the patient stabilized, it was decided to abandon extraction of the rv lead.Attempts were made to unlock the lld ez from the rv lead, but were unsuccessful; therefore, the rv lead/lld were cut and capped and remained in the patient (mdr #3007284006-2024-00031).The patient survived the procedure.This report captures the lld ez providing traction within the ra lead when the perforation occurred, requiring intervention.There was no alleged malfunction of the lld ez within the ra lead during the procedure.
 
Manufacturer Narrative
D4): device lot number, expiration date unk.Partial udi populated.H3): the device was discarded, thus no investigation could be completed.H6): cardiac perforation is a known risk of complication with use of the lld.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LLD EZ LEAD LOCKING DEVICE
Type of Device
STYLET, CATHETER
Manufacturer (Section D)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key18647494
MDR Text Key334616548
Report Number3007284006-2024-00030
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number518-062
Device Catalogue Number518-062
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MEDTRONIC 4194 LV LEAD; MEDTRONIC 5944 RA PACING LEAD; MEDTRONIC 6944 RV PACING LEAD; SPECTRANETICS 14F GLIDELIGHT LASER SHEATH; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS LLD EZ IN RV LEAD; SPECTRANETICS MEDIUM VISISHEATH DILATOR SHEATH
Patient Outcome(s) Required Intervention; Life Threatening;
Patient Age61 YR
Patient SexMale
Patient Weight73 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-