Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/06/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).G2 foreign source: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a knee procedure, a hair was found in the sterile packaging of the tibial insert.Another tibial insert was used to complete the procedure without incident.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Visual evaluation of the provided photo found the packaging was previously opened.A hair-like fiber was found in the sterile packaging.As the product was previously opened, the source of the debris cannot be confirmed.Device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for the reported part and part lot combination.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was previously opened.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|