Catalog Number CDS0706-NTW
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Event Description
It was reported that a mitraclip procedure was performed to treat functional mitral regurgitation (mr) grade 4.An ntw mitraclip (lot: 30512a2065) was chosen for implantation.Preparation was performed per ifu without incident.While in the anatomy during gripper orientation, one of the grippers did not lower.Troubleshooting was performed but the issue persisted.The device was removed from the anatomy and the issue still persisted.Replacement ntw was used to complete the procedure, reducing mr to grade 1-2.There was no patient consequences or clinically significant delay.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information /na.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and without the device to analyze, the cause of the reported single gripper actuation issue was unable to be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18651183
MDR Text Key334661541
Report Number2135147-2024-00519
Device Sequence Number1
Product Code NKM
UDI-Device Identifier05415067037435
UDI-Public(01)05415067037435(17)240516(10)30512A2065
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/16/2024
Device Catalogue NumberCDS0706-NTW
Device Lot Number30512A2065
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/16/2024
Initial Date FDA Received02/06/2024
Supplement Dates Manufacturer Received03/22/2024
Supplement Dates FDA Received03/26/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Age83 YR
-
-