Model Number BF-PE2 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/15/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device has not yet been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.There was no patient infection.According to the customer, there are no concerns or deficiencies in reprocessing.The customer did not provide detailed information on the cleaning, disinfection, and sterilization (cds).Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
|
Event Description
|
It was reported that, during reprocessing, the bronchofiberscope tested positive for 9 colony forming units (cfus) of staphylococcus aureus.After customer observed the positive culture, the device was quarantined.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer completed cleaning disinfection and sterilization (cds) checklist, results of third-party testing, the device evaluation and the lms final investigation.The lm reviewed the customer provided the cds processes where information to judge deviation from the instructions for use (ifu) was not identified.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: 2cfu.Bacterial identification: bacillaceae.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|