• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSERT-IQ¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSERT-IQ¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION) Back to Search Results
Model Number DM5500
Device Problems Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Patient Problem Unspecified Infection (1930)
Event Date 01/22/2024
Event Type  Injury  
Manufacturer Narrative
Further information was requested but not received.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
 
Event Description
It was reported that the patient presented with an unspecified infection.The insertable cardiac monitor (icm) was explanted.
 
Manufacturer Narrative
Correction: upon review the insertable cardiac monitor (icm), manufacturer report 2017865-2024-03952, should not have been submitted as a medical device report (mdr) as infection that is not related to the implant site or implanted device is a non-reportable complaint, as the infection occurred as the result of another factor or source.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSERT-IQ¿ ICM
Type of Device
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18656379
MDR Text Key334720946
Report Number2017865-2024-03952
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K230286
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberDM5500
Device Lot NumberS000091050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/07/2024
Supplement Dates Manufacturer Received02/08/2024
Supplement Dates FDA Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age45 YR
Patient SexFemale
-
-