Qn# (b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The photo shows a peel-away sheath/dilator assembly with signs of crimping on the peel-away sheath body and tip.Therefore, the complaint of peel-away sheath body damaged was able to be confirmed by the photo, however, a complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not use excessive force when introducing guidewire, peel-away sheath over tissue dilator, or tissue dilator as this can lead to vessel perforation and bleeding." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|