Model Number CLV-190
Device Problems Leak/Splash (1354); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned as part of the annual inspection and found the device was leaking due to the outlet connector and tube being worn.Additionally, the evaluation found the following: a damaged tank holder and the output socket was worn out.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the investigation, the phenomenon was attributed to stress of repeated use, external factors, or handling.The root cause of this event was unable to be identified.The root cause of this event was unable to be identified.Olympus will continue to monitor field performance for this device.
 
Event Description
A user facility returned the olympus asset, light source, to the olympus service center for annual inspection.Upon inspection and testing of the returned device, it was observed that traces of water were in the scope socket and inside the tube from the scope socket.This report is being submitted for the reportable malfunction found during evaluation.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18659048
MDR Text Key335712904
Report Number3002808148-2024-01175
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-