MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number 7122Q/58
Device Problems Difficult to Fold, Unfold or Collapse (1254); Retraction Problem (1536); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2024
Event Type  Injury  
Event Description
It was reported the patient presented with a right ventricular lead which became dislodged.The physician attempted to reposition the lead; however, the helix could not be moved.The lead was explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
Further information was requested but not received.
 
Manufacturer Narrative
The reported events were lead dislodgement and helix mechanism issue.A complete lead was returned in one piece for analysis.The reported event of helix mechanism issue was confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual inspection found the helix fully extended and clogged with blood.X-ray examination found over-torqued of the inner coil at the connector region and stretched helix at the helix region consistent with procedural damage.The helix mechanism/extension length test could not be performed due to the helix being damaged and clogged with blood.The cause of helix mechanism issue was isolated to helix being damaged, helix clogged with blood, and over-torque of the inner coil.
 
Event Description
Additional information received indicates the lead dislodgement was identified through diagnostic imaging.The helix could not be retracted or extended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Type of Device
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18659686
MDR Text Key334766510
Report Number2017865-2024-03977
Device Sequence Number1
Product Code NVY
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7122Q/58
Device Lot NumberA000146565
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/20/2024
Initial Date FDA Received02/07/2024
Supplement Dates Manufacturer Received02/14/2024
03/25/2024
Supplement Dates FDA Received02/16/2024
03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-