Model Number FX576T
Device Problem Mechanical Problem (1384)
Patient Problem Hydrocephalus (3272)
Event Date 01/24/2024
Event Type  Injury  
Event Description
It was reported that a progav 2.0 (#fx576t) was implanted during a procedure performed on (b)(6) 2022.According to the complainant, the valve could not be adjusted.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 86 years.Weight: 70 kg.Height: 175 cm.
 
Manufacturer Narrative
Investigation: visual inspection: during the investigation, signs on the outer housing of the valve, but no significant deformations or damage was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: the results show that the valve operates within the accepted tolerances in the horizontal position.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational; however, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection : after dismantling of the valve, deposits were found in progav 2.0.Result : based on our investigation results, we can determine adjustment difficulties in the valve.The determined deposits can be named as the cause for the adjustment difficulties.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18663313
MDR Text Key334808303
Report Number3004721439-2024-00025
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFX576T
Device Catalogue NumberFX576T
Device Lot Number20061337
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/30/2024
Initial Date FDA Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age86 YR
Patient Weight70 KG
-
-