|
Model Number 80617 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/16/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that catheter was punctured.A renegade stc 18 was selected for use in a severely calcified vessel.During the procedure, upon insertion of the guidewire, it was noted that the catheter was punctured.The procedure was completed with another of the same device.No patient complications were reported, and the patient was fine post procedure.
|
|
Event Description
|
It was reported that catheter was punctured.A renegade stc 18 was selected for use in a severely calcified vessel.During the procedure, upon insertion of the guidewire, it was noted that the catheter was punctured.The procedure was completed with another of the same device.No patient complications were reported, and the patient was fine post procedure.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: the device was returned for evaluation.The device was visually and microscopically inspected for damage.The renegade device showed multiple bends and kinks throughout the catheter shaft.A hole in the catheter shaft was confirmed located 98cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Search Alerts/Recalls
|
|
|